These measures are very effective for extending the life of low-speed refrigerated centrifuges

Low speed refrigerated centrifuge is a multi-purpose high-speed large capacity refrigerated centrifuge, with advanced technology intelligent centrifuge. Widely used in clinical medicine, biochemistry, genetic engineering, immunology and other fields. It is an instrument used for centrifugal separation in hospitals, scientific research units and universities at all levels.

These measures are very effective for extending the life of low-speed refrigerated centrifuges

Low-speed refrigerated centrifuge life extension measures:
1. After centrifugation, dry the water in the centrifugal chamber, and apply a little neutral lubrication grease on the cone of the motor spindle every week to prevent corrosion of the rotating shaft. If you do not need a large capacity refrigerated centrifuge for a long time, the rotor should be removed, wiped and placed in a dry place to prevent rust.

2, the main power plug should be removed when the instrument is not used for a long time or maintenance. Otherwise, the instrument will be charged, especially when maintenance is prone to safety accidents.

3, in order to protect the refrigeration compressor, the interval between the instrument and the power is greater than 3 minutes, otherwise the compressor will be damaged.

4. When the rotor is not in use, it should be removed from the centrifugal chamber, cleaned and dried with neutral detergent in time to prevent chemical corrosion, and stored in a dry and ventilated place. It is not allowed to scrub the rotor with non-neutral detergent, and it is not allowed to dry the rotor with hot air. The center hole of the rotor should be protected by a little grease.

5, in order to ensure the freezing effect, when the ambient temperature is higher than 30 ° C, the rotor and centrifugal chamber should be pre-cooled, the rotor should also reduce the speed of 15% operation.

6, the centrifugal tube should be updated regularly, it is strictly prohibited to use the centrifugal tube on the verge of rupture.

7, before each use should pay attention to check the rotor for corrosion points and fine cracks, prohibit the use of corroded or cracked rotors, the use of more than the shelf life of the rotor, to ensure personal safety.

8, large capacity refrigerated centrifuge rotor use must be confirmed to set the rotor number is correct. If the rotor number is set incorrectly. It will cause the rotor to overspeed or not achieve the desired centrifugal effect. In particular, the use of excessive speed may lead to the fatal accident of the rotor explosion, which must not be negligent.